19. EXCITATION OUTPUT & POWER SUPPLY
Three isolated transducer excitation output levels are available from the power supply board.
These are selectable via jumpers b, c, d, e, f in the upper right of the board, as illustrated.
In addition, the board provides three jumper positions for special features. The same jumper
locations apply to the universal power supply (85-264 Vac) and to the low voltage power
supply (12-32 Vac or 10-48 Vdc).
Excitation output Jumper locations
5 Vdc ±5%, 100 mA max
10 Vdc ±5%, 120 mA max
24 Vdc ±5%, 50 mA max
b, d, e
b, d, f
c
SELECTION OF OTHER JUMPERS
Jumper a
- Front panel menu lockout, locked when installed. (See Section 9)
Jumper g
- Provides +5V power output at P1-4 when installed.
Jumper h
- Connects "Control Input 2" to P1-4 when installed.